Water in Turbidimeter #1

Open
smp339 opened this issue Oct 23, 2019 · 1 comment
Open

Water in Turbidimeter #1

smp339 opened this issue Oct 23, 2019 · 1 comment

Comments

@smp339
Copy link
Contributor

smp339 commented Oct 23, 2019

Turbidimeter wasn't working, so we opened it and found water sitting in it. We're trying to dry it, but we might need a new turbidimeter. We are also going to check the remaining turbidimeters for the Unit Processes team for water. @lainey-reed @monroews @adalian @TysonNReed @jenaroz @leb252 @rbruzual

@monroews
Copy link

Can you pull all of the sample vials out of the turbidimeters and hang the vials in front of the meters so that if there is a leak anywhere it is apparent?

The sample vials should stay out of the meters until the flow of water has been tested and everything has been proven to be free of leaks.

Make sure the turbidimeter that was wet is unplugged and open it up so it can dry.