bead normalising when less beads in some tubes #55

Open
stefk7 opened this issue Oct 2, 2023 · 1 comment
Open

bead normalising when less beads in some tubes #55

stefk7 opened this issue Oct 2, 2023 · 1 comment

Comments

@stefk7
Copy link

stefk7 commented Oct 2, 2023

Hi, I had an issue with the beads in the first batch of my mass cytometry experiment. The beads were sticking to the plastic tube and therefore not enough were in solution of my sample to normalise my data properly and this is why I am using premessa to bead normalise my data. The bead normalised graphs premessa is producing look different in batch 1 to the ones being produced for my other 2 batches (which had enough beads). The slope of the graphs Is still straightening out after normalising but they look different, does anyone know if they are still being normalised properly?

premessa graphs
@SamGG
Copy link

SamGG commented Oct 2, 2023

Hi,
No answer for you. You may look at a well expressed marker such as CD3 or CD45 versus time and check if the signal is stable or still decreasing.
The best would be to clearly label the batches on the screenshot and post your question on cytoforum where you will get help from the whole community.
Issues in github are more dedicated to software questions (installations, bugs...).
Best,
Samuel

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment