Merge pull request #40 from MOchiara/patch-4
Browse files Browse the repository at this point in the history
Update chla_post_recovery.md
  • Loading branch information
MOchiara authored Aug 15, 2024
2 parents 38e0d2f + 5fff6ab commit 2835db6
Show file tree
Hide file tree
Showing 2 changed files with 21 additions and 7 deletions.
12 changes: 9 additions & 3 deletions sections/chla_mission_execution.md
Original file line number Diff line number Diff line change
Expand Up @@ -6,13 +6,19 @@ This section covers the activities of those deploying and recovering the gliders
## Deployment
...


## In-situ reference samples
...

## Piloting
(deployment_small_boat)=
## In situ intercomparison during deployment/recovery from a small boat
...

(deployment_ship_CTD)=
## Calibration during deployment/recovery from a ship with a CTD rosette equipped with a calibrated oxygen sensor
...

### Gather data to help correct for sensor response time
## Piloting
...

### Gather data to correct for sensor drift
...
16 changes: 12 additions & 4 deletions sections/chla_post_recovery.md
Original file line number Diff line number Diff line change
Expand Up @@ -4,13 +4,21 @@
At first users should report that their mission is over to support(at)oceanops.org

## Biofouling assessment
...
Pictures should be taken for biofuiling assessment.
Do we have good photos of biofouling?


## Storage and cleaning
...
After recovery and once all caibrations are done, clean the sensor and remove any biofouling. The following protocol is recommended by SeaBird for the Wetlabs EcoPUCK:

1. Remove the protective copper plate, if present.
2. To clean the optics head use soap and water with a non abrasive pad.
3. If necessary, a wooden scraper can be used to push the material from the optics area.
4. Rinse very well with clean water and dry carefully.
5. Do not use any strong solvent (ie. acetone or MEK) which could ruin the epoxy head of the sensor

## Lab calibration
...
When the glider is recovered, tape and check for dark count stability or use the seabird calibration stick?

## Field calibration
...
If you recover the glider from a small boat or a research vessel follow protocols described in {numref}`deployment_small_boat` and {numref}`deployment_ship_CTD`, respectively.