Conversation
Codecov ReportAll modified and coverable lines are covered by tests ✅
Additional details and impacted files@@ Coverage Diff @@
## main #372 +/- ##
=====================================
Coverage 1.75% 1.75%
=====================================
Files 20 20
Lines 4841 4841
=====================================
Hits 85 85
Misses 4756 4756 ☔ View full report in Codecov by Sentry. 🚀 New features to boost your workflow:
|
|
Solves mehta-lab/waveorder#264 |
talonchandler
left a comment
There was a problem hiding this comment.
Thanks for the first pass, @edyoshikun.
| - Connect the external power supply to the triggerscope | ||
| - Flip on the switch | ||
|
|
||
| ### Remove/Check the Meadowlark device is not in config file |
There was a problem hiding this comment.
Is this necessary? I think there's a Triggerscope + Meadowlark config file on automaton, but I haven't tested it thoroughly.
docs/triggerscope-set-up.md
Outdated
| --- | ||
| ## Triggerscope | ||
| The Triggerscope is ... | ||
| The triggerscope replaces the Meadowlark control box and allows one to drive the LCs through external analog inputs. |
There was a problem hiding this comment.
The Meadowlark DS5020 is still necessary, right? AFAIK it's converting the input DC voltages to an unbiased 20 kHz square wave.
There was a problem hiding this comment.
yes, you are correct.
| Launch Micromanager, open `Devices > Hardware Configuration Wizard...`, and add the `Triggerscope Hub` device to your configuration. | ||
|
|
||
| Confirm your installation by opening Devices > Device Property Browser... and confirming that `Triggerscope DAC` properties appear. | ||
|
|
There was a problem hiding this comment.
I'd benefit from details about what the "Sequencing" settings do.
|
Preview page for your plugin is ready here: |
|
closing in favor of (mehta-lab/waveorder#442) |
This PR adds documentation on how to setup the triggerscope to drive the LCs.